MODEL 1000 SEP 2005
INSTALLATION AND SETUP
3-89
(b) Ensure that the power source leads are properly connected to the HOT, NEUTRAL,
and GROUND terminals. (Generally, the color convention for these wires is HOT -
black, NEUTRAL - white, and GROUND - green.)
(c) Make power line splices and conduit seals that comply with applicable wiring
requirements (for hazardous environments).
EQUIPMENT DAMAGE OR PERSONAL INJURY
Do not apply AC electrical power to the Analyzer or the GC Controller until all electrical
power, interconnection, and external signal connections have been verified, and proper
grounds have been made.
Failure to properly connect the GC unit may result in serious equipment damage or personal
injury.
(4) If necessary, connect the Controller's chassis ground to an external copper ground rod (at
remote locations). See Section 3.1.4, this manual, regarding electrical and signal ground.