MODEL 1000 SEP 2005
EQUIPMENT DESCRIPTION
2-9
2.2.5 Analyzer Specifications
a. Power Requirements: 120 volts, alternating current (VAC), +10/-15VAC; 50 to 60 Hertz
(Hz); single phase; 10 amperes (A) (maximum during warmup) (additional 10 amps if unit
has a Sample System Oven).
b. Ambient Temperature Range: -18 degrees Celsius (°C) to +55°C (0°F to +130°F)
c. Humidity: 0 to 95 percent relative humidity, noncondensing
d. Vibration: Designed for mounting on process piping or other field structures subject to
normal process vibrations
e. National Electrical Code (NEC) Area Classification: Suitable for NEC Class 1, Division 1,
Group D
f. Frame Size (approximately):
- Height: 78 inches (198 centimeters [cm])
- Width: 24 inches (61 cm) maximum
- Depth: 24 inches (61 cm)
g. Weight: Approximately 125 pounds (56.8 kilograms [kg]), including mounting hardware.
h. Sample Requirements:
1. Fluid Phase - Vapor
2. Pressure - 15 to 30 psig, regulated to ±10 percent
3. Flow Rate - 50 cc/min, typical
i. Analyzer Output Signal: Four different gain channels to provide a 4 to 20 mA signal to the
Controller.
j. Transient Over Voltages Category:
Installation Category (Over Voltage Category II)
k. Cleaning requirements are restricted to the 6-port valve (refer to Section 5.5.5.1 Valve
Cleaning, in the manual).