This Appendix provides a description of the Carrier Manifold that permits connection of two carrier
gas bottles, or cylinders, to a Gas Chromatograph (GC) System (part number for the manifold is 3-
5000-050). The benefits of this manifold are as follows:
- When one bottle is nearly empty (i.e., 100 psig remaining), the other bottle becomes the
primary supply.
- Each bottle can be disconnected for refilling without interrupting GC operation.
The illustration and information in this Appendix are from drawing AE-10098.
This Appendix is organized as follows:
Illustration ...................................................... See Section B.1
Installation and Line Purging ....................................... See Section B.2
Replacing Carrier Cylinder ........................................ See Section B.3