EasyManua.ls Logo

Emerson Daniel Danalyzer 1000 - Discrete (Digital) I;O Wiring

Emerson Daniel Danalyzer 1000
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODEL 1000 SEP 2005
INSTALLATION AND SETUP
3-83
3.4.6 Discrete (Digital) I/O Wiring
4
Install Analyzer Wiring
5
Install Analyzer Sample & Gas Lines
6
Install GC Controller Wiring
»
7
Perform Leak Checks
To connect digital signal input/output lines to the GC Controller, follow these steps:
(1) Access the GC Controller's Terminal Board for Field Wiring (TB).
(2) Route digital I/O lines appropriately. Refer to drawing DE-20782.
(3) Make digital I/O connections to the GC Controller TB. There are connections for five digital
input and five digital output lines, as follows:
(a) Digital Inputs - GC Controller TB, connection ports "J7" and "J9"
Digital Outputs - GC Controller TB, connection port "J8"
(b) For specific terminal, or pin number assignments, see drawing DE-20782, GC
Controller drawings addendum, this manual.

Table of Contents