EasyManua.ls Logo

Emerson Daniel Danalyzer 1000 - Baseline Drift

Emerson Daniel Danalyzer 1000
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODEL 1000 SEP 2005
MAINTENANCE
5-15
5.5.4.3 Baseline Drift
To ensure that the baseline is not drifting, compare the baseline upsets caused by valve actuations
with those of the chromatogram provided with the Operational Parameters Sheet. Ensure that no
evidence of component elutions is present when no sample is being injected.
If differences exist between the two chromatograms, the problem may be due to one or more of the
following:
1. Programming of events
2. Contamination of valve sealing diaphragms by foreign matter
3. Improperly adjusted flows
4. Leaks in the carrier system
5. Column deterioration due to liquid contamination from a sample
6. Misidentifying peaks
A noisy baseline can be caused by carrier leaks, an electronic failure in the Preamplifier, a faulty
power supply, or defective thermistors in the detector. If the baseline is still noisy after correcting
for leaks, perform the Detector Bridge Balance procedure before replacing the detector thermistors
or the preamplifier board.

Table of Contents

Related product manuals