MODEL 1000 SEP 2005
INSTALLATION AND SETUP
3-95
Figure 3-39. Analyzer Valve Driver Board (in upper explosion-proof enclosure)
(13) Regulate the flow through the rotameter to approximately 50 cubic centimeters per minute
(cc/min) by adjusting the metering valve on the Flow Panel.
(14) Prepare for normal operation, as follows:
(a) On the Valve Driver Board, upper enclosure, reset the Stream switch "S2" to AUTO
(if Stream 2 will be used for calibration gas).
(b) Close upper enclosure.