SEP 2005 MODEL 1000
OPERATION FROM LOCAL KEYPAD AND DISPLAY
4-10
To halt ongoing analysis runs, proceed as follows:
(1) In the GC Control submenu, press the 4 key to select "Halt."
(2) At the "Halt - Write Changes" prompt, press the ENTER key. (Or, to return to the GC
Control submenu without halting analysis runs, press the ESC key.)
When analysis runs are halted at the end of the current analysis cycle, the green LED indicator will
no longer be illuminated.
If you choose "Stop Now", instead of "Halt", it forces the system to an immediate idle. This
may allow components to continue to elute from the columns. Therefore, it is recommended
you choose "Halt" instead of "Stop Now." A "Halt" command stops analysis runs after the
current run is finished.