MODEL 1000 SEP 2005
MAINTENANCE
5-53
Table 5-7. Function Code Dictionary in Hexadecimal
INPUT (hexadecimal) VALVE = SET (ON) VALVE = RESET (OFF)
0 Analyzer Valve (AV)-4
1 AV-4
2 AV-5
3 AV-5
4 AV-1
5 AV-1
6 AV-2
7 AV-2
8 Stream Valve (SV)-1 SV-2, -3, -4, -5
9 SV-2 SV-1, -3, -4, -5
A SV-3 SV-1, -2, -4, -5
B SV-4 SV-1, -2, -3, -5
C SV-5 SV-1, -2, -3, -4
D AV-3
E AV-3
F
A measurement of 5 (five) volts at any of the function code terminals indicates a binary zero
(0). A measurement of 0 (zero) volts indicates a binary one (1).