MODEL 1000 SEP 2005
SETTING SOLENOID PURGE FLOWS
E-3
(8) When flows for all streams, except stream 1, are set, manually turn stream 1 OFF at the
Analyzer Valve Driver Board.
(a) Now, manually turn stream 2 ON at the Analyzer Valve Driver Board.
(b) Next, open the solenoid purge valve associated with stream 1 to add 200 ccm to the
current flow (as measured at the solenoid purge flow indicator).
(9) When all flows have been set, return all stream switches to AUTO positions (at the Analyzer
Valve Driver Board).
(10) After you have set all stream switches to AUTO positions, close the upper purged enclosure
(that houses the Analyzer Valve Driver Board), and commence normal analysis.