SEP 2005 MODEL 1000
2350 to 2350A CPU RETROFIT INSTRUCTIONS
F-4
(14) Ensure that DIP Switches 6, 7, and 8 of S1 are all in the “OFF” position. If the 2350A is
powered up with S8 in the “ON” position, it will delete the current application (commonly
called a “Cold Start”).
(15) To inspect or change the GC Controller's COM ID setup at the GC Controller site, locate the
DIP switch as described in the following steps.
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Before removing the unit cover from the GC Controller, make certain the power supply
switch is OFF and the AC power cord is disconnected. Observe all safety precautions
when you are working in a hazardous environment.
Failure to observe all safety precautions could result in serious injury or death.