MODEL 1000 SEP 2005
INSTALLATION AND SETUP
3-33
To inspect or change the GC Controller's COM ID setup, follow these steps:
(1) At the GC Controller site, locate the DIP switch as described in the following steps.
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
Before removing the unit cover from the GC Controller, make certain the power supply
switch is OFF and the AC power cord is disconnected. Observe all safety precautions
when you are working in a hazardous environment.
Failure to observe all safety precautions could result in serious injury or death.
(2) For integrally-mounted Controllers, use a flat head screw driver to remove the access panel
on the right side of the card cage assembly (see Figure 3-4).
Figure 3-4. Right Side View of Rack Mount and Panel Mount Controllers