For IS parameters, see Product certifications.
External Supply Voltage:
• Passive Non-IS: 7.2 - 35 Vdc
• Passive IS: 8.7 – 30 Vdc
Maximum Output Voltage (open loop):
• Active Non-IS: 24 Vdc
• Active IS: 23 Vdc
HART master:
• Maximum 5 HART Slave Devices (Passive)
• Maximum 3 HART Slave Devices (Active)
Figure A-1: Loop Resistance: Passive Non-IS Analog Input
%
$
$
A. Loop resistance [Ω]
B. External power supply voltage [V]
Max Loop Resistance
(14)
@ 23 mA = 43.4 * (External Power Supply Voltage -7.2) [Ω]
(14) Any sense resistance must be subtracted from calculated max loop resistance to receive the maximum cable resistance.