EasyManua.ls Logo

Emerson Rosemount 2410 - Page 184

Emerson Rosemount 2410
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7. In the TankMaster WinSetup workspace, right-click the ATD device icon and choose
the Properties option.
8. Select the Analog Input tab.
9. Ensure that the Enable check box is selected.
10. In the Value Unit list, select a suitable unit to match the input source.
11. Under Value Range, enter the parameter values that correspond to the analog
input values 4 and 20 mA.
You can specify any value as long as the 20 mA value is higher than the 4 mA value.
If the measured value goes outside the range values, the analog input enters alarm
mode.
12. Optional: Click the Advanced button in case you would like to configure Filter
Factor and Current Limits.
Advanced configuration Reference Manual
December 2019 00809-0100-2410
184 Reference Manual

Table of Contents

Other manuals for Emerson Rosemount 2410

Related product manuals