EasyManua.ls Logo

Emerson Rosemount 2410 - Page 82

Emerson Rosemount 2410
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3. Right-click and select Live List.
Figure 6-1: The Device Live List Window
Device live list window
The Rosemount 2410 Tank Hub Device Live List window shows the following information:
Table 6-1: Device Live List Description
Item Description
Device Type Examples of supported devices: Rosemount 5900S, 2410, 2240S, 2230,
5300, 5408, 848T, and 3051S.
For unknown devices the device type number is shown.
Device ID (Unit ID) A unique code that identifies a particular device. You can choose to view
the device ID in decimal or hex format depending on what format is
supported by the device.
Manufact Id Identifies the manufacturer.
Device No Index used to identify devices by the FF stack.
FF Address FOUNDATION
Fieldbus address used for communication on the Tankbus.
Handled Bit 0 of the Live List Status input register which indicates the current
Tankbus communication status of the device.
Connected “No” means that the device has been disconnected from the Tankbus.
Configured “Yes” indicates that the device is configured in the Rosemount 2410
tank database, i.e. the device is mapped to a particular tank.
Opened Bit 1 of the Live List Status input register which indicates the current
Tankbus communication status of the device.
Auto Mode “Yes” in normal operation. “No” indicates that the device is in Out of
Service mode.
Tag A removable tag provided with the device lets you identify the device to
a physical location. This field shows the Tag number of the device (when
available).