6.2.14 Logging measurement data using TankMaster
The Rosemount 2410 supports logging of diagnostic registers. This function is useful for
verifying that the gauge works properly. The logging function can be accessed by using
the Rosemount TankMaster WinSetup program.
Procedure
1. Start the Rosemount TankMaster WinSetup program.
2. In the TankMaster WinSetup workspace window, select the device icon.
3. Right-click and select Logging.
4. Select Manual or Automatic mode.
Option
Description
Manual Manual mode lets you start logging at any time.
Logging will proceed until it is stopped by clicking the
Stop button.
Automatic In Automatic mode you have to specify a Start and
Stop time. Logging will proceed until the stop date
and time is reached.
The resulting log file will not exceed the size specified by the Max File Size
parameter. When the number of log files has reached the Max Log Files value,
TankMaster starts replacing the contents of existing log files.
Reference Manual Service and troubleshooting
00809-0100-2410 December 2019