Emerson Process Management GmbH & Co. OHG 4-7
4
Maintenance
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
03/2012
4.2 Replacement of Parts
Specically for CSA-C/US approved analyzers
Metric-to-NPT adapters
Any CSA-C/US certied ameproof type
Male thread: M20 x 1,5
Female thread:
1
2“ or
3
4“ per request
e.g.
- AD-U series by Redapt Ltd.
3
4“ version spare part # ETC01877
Cable entries stopping plugs
Any CSA-C/US certifed ameproof type
Male thread: M20 x 1,5
e.g.
- PD-U by Redapt Ltd.
Spare part # ETC01878
For ATEX and CSA analyzers
Flame arrestors
Type: FA 01, FA 02, FA 03
by EMERSON PROCESS MANAGEMENT
FA 01: Spare part # 42716625
FA 02: Spare part # 42716625-V1
FA 03: Spare part # 42716625-V2
Enclosure screws
Material: stainless steel A2-70
Thread: M16 x 45 ISO 4762 A2
Modied to drawing 4.271-6945
by EMERSON PROCESS MANAGEMENT
Spare part # 42716945
Internal Battery
Type: CR2032 by Varta;
Lithium button cell
3Vdc; 230 mAh

Table of Contents

Other manuals for Emerson Rosemount Xstream Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount Xstream Series and is the answer not in the manual?

Emerson Rosemount Xstream Series Specifications

General IconGeneral
BrandEmerson
ModelRosemount Xstream Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals