EasyManua.ls Logo

Endress+Hauser analytikjena PlasmaQuant 9100 - Figure 25 Autosampler Page, Dilute Tab

Endress+Hauser analytikjena PlasmaQuant 9100
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation and commissioning PlasmaQuant 9100 (Elite)
40
} Connect the switching valve to the dilution system ("ICP/ASXpress" connection) via
hose connection 2 ("Autosampler") on the 6-port valve.
} Connect the switching valve with the sample hose of the emission spectrometer via
hose connection 5 ("Nebulizer").
Software support during instal-
lation and commissioning
Control of the sampler and the dilution system is integrated into the ASpect PQ soft-
ware.
Figure25Autosampler page, Dilute tab
For commissioning and maintenance of the sampler and the dilution system, use the
following software commands in the
Dilute
tab in the
Service
area:
Time Measure Software command
¡ Installation of the di-
lution system
¡ During syringe re-
placement
Inserting the syringe in
the syringe pump of the
dilution system.
Move syringe plunger to
½ position for this.
Move syringe to removal
position
¡ Commissioning of the
dilution system after
installation or main-
tenance
¡ After changing the
diluent
¡ For cleaning after use
of a highly acidic or
alkaline purging solu-
tion or an organic
solvent
Purging the syringe
pump, hoses and vortexer
mixing vessel with purg-
ing solution.
Removing air bubbles
from the hoses.
Prime syringe and vor-
texer
¡ After weekly cleaning
and after replace-
ment of the 6-port
valve
Initializing the switching
valve.
The control electronics
set the internal valve
stops and determine the
correct stroke.
Re-home ASXpress+ after
disassembling for clean-
ing

Table of Contents

Related product manuals