Transport and storage PlasmaQuant 9100 (Elite)
78
8 Transport and storage
8.1 Preparing the device for transport
} Switch on the device and start the control software.
} Remove the cooling water from the system:
– Start the wizard for replacing the cooling water in the control software.
– Drain the cooling water and exit the wizard (→ "Maintenance of the recirculating
chiller: Replacing the cooling water" 76).
} Switch off the device. Exit the control software and switch off the PC.
} Disassemble the torch, spray chamber and nebulizer and pack them.
} Remove the cover plate in front of the connections on the left rear side of the device.
} Disconnect all electrical connection cables from the device, the PC and the sampler.
} Disconnect all cooling water hoses from the device.
} Place an absorbent sheet underneath the connections to catch dripping liquid. Press
the ring on the quick-release connector inwards and pull the hose from the connec-
tion for this.
} Disconnect the argon hose from the device. On the quick-release connector on the
rear left side of the device, press the blue ring inwards and pull out the hose.
} Unplug the interface cables of the electrical components (sampler, PC) from the con-
nections on the connection panel on the left rear side of the device.
} Reattach the cover plate in front of the connections on the left rear side of the de-
vice.
} Screw in the four transport handles as far as possible.
} Pack the device in the original packaging.
8.2 Moving the device in the laboratory
CAUTION
Risk of injury due to falling device
¡ Screw the four transport handles into the device as far as possible. Only this allows
you to grip and carry the device safely.
Observe the following when moving the device within the laboratory:
¡ Insufficiently secured components pose a risk of injury!
Before moving the device, remove all loose parts and disconnect all connections
from the device.
¡ For safety reasons, four persons are required to transport the device, one person at
each corner of the device.
¡ Firmly grip the device with both hands by the transport handles. Lift the device si-
multaneously.
¡ Observe the guide values and adhere to the legally mandated limits for lifting and
carrying loads without auxiliary means.
¡ Observe the installation conditions at the new location.