PlasmaQuant 9100 (Elite) Maintenance and care
77
ð Required tools: 10L distilled/deionized water, cooling water additive set for recircu-
lating chiller, suitable glass, plastic or stainless steel container for mixing the cooling
water, bucket to collect the drained coolant
} Dissolve the contents of both bottles of the cooling water additive set (biocide and
corrosion protection) on 10L of water.
} Start the wizard to replace the cooling water in the ASpect PQ control software. To
do this, select the
Extras
|
Maintenance
menu item and click on the
[Change]
button.
} Follow the instructions in the wizard:
Switch off the recirculating chiller.
On the recirculating chiller, disconnect the cooling water return flow connection
and hold the hose in the collection vessel (bucket).
Switch the recirculating chiller back on and allow it to run until the flow of cooling
water ends and only mist comes out.
Reconnect the hose to the cooling water return flow connection on the recirculat-
ing chiller.
Unscrew the sealing cap from the filling opening on the tank and insert the fun-
nel.
Pour the coolant into the tank until the MAX level mark is reached.
Switch on the recirculating chiller and observe the level indicator. The level drops
when the pump is operating.
Slowly continue to fill the tank with the coolant until the level stabilizes slightly
below the MAX mark.
Remove the funnel and seal the filling opening with the screw cap.
Confirm the closing step in the wizard.
} Wait for the message in the wizard indicating that the coolant has been replaced.
} Exit the wizard.