Operation PlasmaQuant 9100 (Elite)
48
When loading a sequence, make sure that the calibration is compatible with the
method.
Analysis lines of the calibration standards must match the calibration in the
method.
After calibration, measure a QC sample in order to verify the correctness of the
calibration.
} Create a sample ID table.
} Starting the measurement:
Start the measurement routine by clicking on or selecting the
Routine
|
Start
sequence
menu item.
In the
Start sequence
window, select or enter a file name for the results file.
ü After selection of the file name, the measurement routine will start according to
the settings made in the method and the sequence.
} When using a sampler, the measurement runs automatically.
When performing sampling manually without a sampler, follow the instructions in
the software on sample supply.