PlasmaQuant 9100 (Elite) Specifications
83
Display types
Emission Counts (ct)
Intensity Counts/second (ct/s)
Concentration 5-digit value range (0.0001 to 99999), unit freely selectable
Signal evaluation
Spectral resolu-
tion
Spectra widths of 20 to 200 pixels
Analytical data
Sample type Liquid
Type of nebulizer Concentric nebulizer
Spray chamber Cyclone chamber
Power supply
Voltage 230 V ±10%
Frequency 50/60 Hz
Typical average power con-
sumption
4500 VA
Maximum current consump-
tion
32 A
Fuse protection (grid side) 32 A
Device fuses
Fuse Type Protected circuit
S1 10 A NFC 10x38 gG AC, 400 V Spectrometer
S2 6 A NFC 10x38 gG AC, 400 V Tube heating generator
S3 25 A NFC 10x38 gG AC, 400 V Generator power supply
unit
Safety circuits
Monitoring ¡ Plasma chamber door closing
¡ Torch working position
¡ Cooling
¡ Argon supply
¡ Plasma (optical monitoring)
Gas supply
Gas Inlet pressure Total consumption
Argon ≥ 4.6
Permissible components:
Oxygen ≤ 3 ppm
Nitrogen ≤ 10 ppm
Hydrocarbons ≤ 0.5 ppm
Moisture ≤ 5 ppm
600 kPa (6 bar) 13 to 21 l/min
Oxygen ≥ 4.5
(as optional auxiliary gas)
600 kPa (6 bar) ≤ 0.04 l/min