Gammapilot FMG50 Mounting
Endress+Hauser 23
Examples
A0037715
A Vertical cylinder; the Gammapilot FMG50 is mounted vertically with the detector head pointing either
downwards or upwards, the gamma radiation is aligned to the measuring range.
B Correct: Gammapilot FMG50 mounted outside the tank insulation
C Incorrect: Gammapilot FMG50 mounted inside the tank insulation
D Conical tank outlet
E Horizontal cylinder
1 Source container
2 Gammapilot FMG50
4.2.4 Installation conditions for point level detection
Conditions
For point level detection, the Gammapilot FMG50 is generally mounted horizontally at the
height of the desired level limit.
Measuring system arrangement
A0018075
A Maximum point level detection
B Minimum point level detection
1 Source container
2 Gammapilot FMG50