Operation Gammapilot FMG50
44 Endress+Hauser
A0039243
 15 Display with Bluetooth module
A flashing Bluetooth symbol indicates that a Bluetooth connection is available
Bluetooth communication with the device is possible with a supply voltage of 14 V or
higher. The background lighting of the display is only guaranteed with a supply
voltage ≥ 16 V. The measurement function is guaranteed as of a terminal voltage of
12 V; Bluetooth communication with the device is not possible with this voltage level,
however.
If the available supply voltage drops below the aforementioned thresholds during
operation, the background lighting switches off first before the Bluetooth function is
switched off in order to guarantee the measurement function. A corresponding
warning message is not displayed. These functions are reactivated when sufficient
power is supplied.
If the available supply voltage was already too low when the device was started these
functions are not activated later on.
Operation via SmartBlue App
1
2
3
A0038833
 16 Operation via SmartBlue App
1 Transmitter power supply unit
2 Smartphone / tablet with SmartBlue App
3 Transmitter with Bluetooth module
6.2.5 Heartbeat Verification/Monitoring
The Heartbeat submenu is only available if operating via FieldCare, DeviceCare or
SmartBlue App. It contains Wizards that are available with the Heartbeat
Verification and Heartbeat Monitoring application packages.