Diagnostics and troubleshooting Gammapilot FMG50
92 Endress+Hauser
L
WARNING
If the incorrect time is set, this falsifies the result of the decay compensation. This could
result in a dangerous failure that cannot be diagnosed in the device.
8.9 Behavior in the event of low terminal voltage
8.9.1 General principles
If the terminal voltage is low, the available energy level may not suffice to make all
functions of the device available. To ensure a reliable measurement function, the following
measures are taken depending on the energy available:
For devices with a display (optional): the background lighting of the display and the
Bluetooth function are disabled
For devices without a display: the total available energy is always available to the
sensor
If the energy does not suffice to reliably guarantee the measurement function, an alarm
F801 "Increase supply voltage" is output and the sensor function is switched off.
8.10 History
8.10.1 Firmware history
Firmware version
01.00.00
Initial software
Valid from: 31 August 2019
01.00.01
SIL functions certified
Display background lighting available
Valid from: 10 February 2020
01.00.02
Certified for overfill protection according to German Water Resources Act (WHG)
Behavior in event of excess radiation improved
Behavior of display in event of low power changed (display lighting and Bluetooth are
reactivated when sufficient power supply is available)
Errors are now shown on the display weighted according to their relevance and no
longer according to when they occur
The Wizards for Heartbeat Verification and SIL proof-testing are now also available via
Bluetooth (SmartBlue App update required)
Bugfixes
Valid from: 1 March 2021
The firmware version can explicitly be ordered via the product structure. In this way it
is possible to ensure compatibility of the firmware version with an existing or planned
system integration.
8.10.2 Hardware history
Hardware version
01.00.00 -> initial hardware
Valid from: 31 August 2019
01.00.01 -> display background lighting available (it may be necessary to update the
display firmware)
Valid from: 10 February 2020