Commissioning Gammapilot FMG50
76 Endress+Hauser
1
2
Alarm
on
off
MaxSW
Zero/I
Span/II
Display
34
A0039285
1 Operating key for empty calibration (function I)
2 Operating key for full calibration (function II)
3 DIP switch for alarm current (SW-defined / Min alarm)
4 DIP switch for locking and unlocking the measuring device
• Empty calibration: press and hold the operating key for empty calibration (I) > 3 s
• Full calibration: press and hold the operating key for full calibration (II) > 3 s
• Background calibration: simultaneously press and hold the operating key for empty
calibration (I) and the operating key for full calibration (II) > 3 s
• Reset to factory defaults: simultaneously press and hold the operating key for empty
calibration (I) and full calibration (II) > 12 s. The LED starts flashing. When the flashing
stops, the device is reset to the factory default settings.
7.4.1 Level basic calibration
Calibration time per calibration: 5 min!
1. Reset
Press both keys > 12 s
2. Start background calibration
Press both keys > 3 s
The green LED is lit for one second and starts flashing at an interval of 2 s
3. Start empty calibration
Press the "Zero / 1" key > 3 s
The green LED is lit for one second and starts flashing at an interval of 2 s
Wait 5 min until the green LED stops flashing
4. Start full calibration
Press the "Span / 2" key > 3 s
The green LED is lit for one second and starts flashing at an interval of 2 s
Wait 5 min until the green LED stops flashing
A reset deletes all calibrations!
7.4.2 Status and power LED
A green LED that signals the status and button activation feedback is provided on the
electronic insert.
Behavior of the LED
• The LED flashes once briefly when the measuring device is started
• When a key is pressed, the LED flashes to confirm the key activation
• When a reset is performed, the LED flashes as long as both keys are pressed and the
reset is not yet active (countdown). The LED stops flashing once the reset is active.
• The LED flashes while calibration is being performed via onsite operation