Gammapilot FMG50 Commissioning
Endress+Hauser 81
2. Enter value pairs in the linearization table. A value pair consists of a percentage value
and the associated pulse rate (pulses per second, cnt/s).
The linearized measured value is shown as a percentage.
The linearization table consists of up to 32 value pairs.
Enter as many value pairs as possible to maximize accuracy.
Setting sensors and channels
Channel 1:
FMG50 density measurement (HART output)
PV: density (kg/m3)
SV: pulse rate (pulses per second, cnt/s)
Channel 2:
FMG50 level measurement (HART output)
PV: level (%)
SV: pulse rate (pulses per second, cnt/s)