Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu)
Endress+Hauser 117
Selection
m
ft
in
mm
Factory setting
m
Tank type
Navigation
Setup ® Tank type
Setup ® Advanced setup ® Interface ® Tank type
Description
Defines the tank type
Selection
Metallic
Bypass/pipe
Non metallic
Mounted outside
Depending on the probe there may be further options.
Factory setting
Depending on the type of probe.
Additional information
For coax probes the "Tank type" is always fixed to "Coax".
Tube diameter
Navigation
Setup ® Tube diameter
Conditions
Only visible if the following conditions are met:
"Tank type" = "Bypass/pipe"
The probe is coated.
Description
Defines the diameter of the bypass or stilling well.
Input range
0 to 9999 mm (0 to 390 inch)
Factory setting
80 mm (3.15 inch)
Tank level
Navigation
Setup ® Tank level
Conditions
Only visible for devices with "interface measurement" application package (product structure:
Feature 540 "Application package", Option EB "Interface measurement").