Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54
180 Endress+Hauser
18.3.5 "Data logging" submenu
Assign channel 1-4
Navigation
Diagnostics ® Data logging ® Assign channel 1
Diagnostics ® Data logging ® Assign channel 2
Diagnostics ® Data logging ® Assign channel 3
Diagnostics ® Data logging ® Assign channel 4
Description
Use this function to assign a process variable to the data logging channel.
A total of 1000 measured values can be logged. This means:
1000 data points if 1 logging channel is used
500 data points if 2 logging channels are used
333 data points if 3 logging channels are used
250 data points if 4 logging channels are used
If the maximum number of data points is reached, the oldest data points in the data log are
cyclically overwritten in such a way that the last 1000, 500, 333 or 250 measured values are
always in the log (ring memory principle).
The log contents are cleared if the option selected is changed.
Auswahl
Off
Level
Distance
Interface
Interface distance
Interface thickness
Current output 1
Measured current
Terminal voltage
Electronics temperature
Absolute echo amplitude
Relative echo amplitude
Absolute interface amplitude
Relative interface amplitude
Absolute EOP amplitude
Signal to noise
Tank noise
EOPshift
Dielectric constant of upper medium
Factory setting
Off
Logging interval
Navigation
Diagnostics ® Data logging ® Logging interval