10 Commissioning via operating menu (On-site
display, FieldCare)
10.1
Installation and function check
Make sure that all final checks have been completed before you start up your measuring point:
•
Checklist "Post-installation check" (® ä 56)
•
Checklist "Post-connection check" (® ä 62)
10.2 Adjust the display contrast
•
O + F (pressed simultaneously): increases the contrast.
•
S + F (pressed simultaneously): decreases the contrast.
10.3 Unlock the device
If the device has been locked, it must be unlocked before the measurement can be configured.
10.3.1 Revoke hardware locking
A0013649
å 26
Mesured value screen of a hardware-locked device
The padlock in the header of the measured value screen indicates that the device is hardware-
locked. In order to unlock the device, shift the locking switch (which is located below the display
module and is marked by "WP") into the "off" position (® ä 66).
10.3.2
Revoke software locking
A0013651
å 27
Input prompt for the access code to unlock software-locked parameters.
Parameters affected by the software lock are marked by a padlock in front of the parameter name.
After pressing F an input prompt appears. Enter the user defined locking code to unlock the
device (® ä 67).