Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu)
Endress+Hauser 173
18.3.2 The "Event logbook" submenu
Filter options
Navigation
Diagnostics ® Event logbook ® Filter options
Description
Use this function to select the category (status signal) whose event messages are displayed in the
events list.
The status signals are categorized according to NAMUR NE 107: F = failure, M =
maintenance request, C = function check, S = out of specification
Options
• All
•
Failure (F)
• Maintenance required (M)
• Function check (C)
• Out of specification (S)
• Information (I)
Factory setting
All
Event list
Navigation
Diagnostics ® Event logbook ® Event list
Description
Use this function to display the history of event messages of the category selected in the Filter
options parameter (® ä 173). A maximum of 20 event messages are displayed in chronological
order. If the advanced HistoROM function is enabled in the device, the event list can contain up
to 100 entries.
The following symbols indicate whether an event has occurred or has ended (status symbols):
• : Event has occurred
• : Event has ended
Information on what is causing the message, and remedy measures, can be viewed via the
symbol on the display.
User interface
• For event messages in category I (status signal): status signal, event number, time event
occurred, event text
•
For event messages in category F, M, C, S (status signal): diagnostics event, status symbol, time
event occurred, event text