EasyManua.ls Logo

Endress+Hauser Levelflex FMP53 - Assign limit

Endress+Hauser Levelflex FMP53
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu)
Endress+Hauser 147
Description
Allocates a DI block to the switch output.
Options
Off
No digital block is allocated.
Digital output 1
Digital output 2
Digital output 3
Digital output 4
Factory setting
Off
Assign limit
Navigation
Setup ® Advanced setup ® Switch output ® Assign limit
Prerequisite
Only visible for Switch output function = Limit.
Description
Defines the variable to be checked for limit transgression and allocates it to the switch output.
Options
Off
No measured variable is allocated.
Level linearized
Distance
Interface linearized
Interface distance
Upper interface thickness
Terminal voltage
Electronic temperature
Relative echo amplitude
Relative interface amplitude
Factory setting
Off
Assign diagnostic level
Navigation
Setup ® Advanced setup ® Switch output ® Assign diagnostic level
Prerequisite
Only visible for Switch output function = Event level.
Description
Defines to which class of diagnostic events the output reacts.
Options
Alarm
Alarm + warning
Warning
Factory setting
Alarm