Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54
186 Endress+Hauser
Navigation
Diagnostics ® Device check ® Level signal
Conditions
Only visible if a device check has been performed.
Description
Displays the result of the device check for the level signal.
Display
Check not done
Check not ok
Check the mounting position of the device and the dielectric constant of the medium.
Check ok
Launch signal
Navigation
Diagnostics ® Device check ® Launch signal
Condition
Only visible if a device check has been performed.
Description
Displays the result of the device check for the launch signal (fiducial).
Display
Check not done
Check not ok
Check the mounting position of the device. In non-metallic vessels use a metal plate or a metal
flange.
Check ok
Interface signal
Navigation
Diagnostics ® Device check ® Interface signal
Condition
Only visible for devices with interface measurement option and if a device check has been
performed.
Description
Displays the result of the device check for the interface signal.
Display
Check not done
Check not ok
Check ok