Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54
186 Endress+Hauser
Navigation
Diagnostics ® Device check ® Level signal
Conditions
Only visible if a device check has been performed.
Description
Displays the result of the device check for the level signal.
Display
• Check not done
•
Check not ok
Check the mounting position of the device and the dielectric constant of the medium.
• Check ok
Launch signal
Navigation
Diagnostics ® Device check ® Launch signal
Condition
Only visible if a device check has been performed.
Description
Displays the result of the device check for the launch signal (fiducial).
Display
• Check not done
•
Check not ok
Check the mounting position of the device. In non-metallic vessels use a metal plate or a metal
flange.
• Check ok
Interface signal
Navigation
Diagnostics ® Device check ® Interface signal
Condition
Only visible for devices with interface measurement option and if a device check has been
performed.
Description
Displays the result of the device check for the interface signal.
Display
• Check not done
•
Check not ok
• Check ok