Levelflex FMP51, FMP52, FMP54 Mounting
Endress+Hauser 43
Bypasses and stilling wells
³ 100
( 4)³
³
³
100
( 4)
Ø
d
Ø
D
Ø
d
25Nm±5Nm
25Nm±5Nm
mm(in)
1 2
3.1
3.2
3
A0012615
1 Mounting in a stilling well
2 Mounting in a bypass
3 Center washer
3.1 Metallic center washer (316L) for level measurement
3.2 Non-metallic center washer (PEEK, PFA) for interface measurement
For information on bypass solutions from Endress+Hauser please contact your Endress
+Hauser sales representative.
Feature 610 - Accessory mounted
Application Option Type of probe
Center washer Pipe
Æ d [mm (in)]
Material
Æ D [mm (in)]
Level measurement OA Rod probe 75 (2,95) 316L DN80/3" to
DN100/4"
OB Rod probe 45 (1,77) 316L DN50/2" to
DN65/2½"
OC Rope probe 75 (2,95) 316L DN80/3" to
DN100/4"
Interface measurement OD Rod probe 48...95
(1,89...3,74)
PEEK
³ 50 mm (2")
OE Rope probe 37 (1,46) PFA
³ 40 mm (1.57")
• Pipe diameter: > 40 mm (1.6") for rod probes
•
Rod probe installation can take place up to a diameter size of 100 mm. In the event of larger
diameters, a coax probe is recommended.
• Side disposals, holes or slits and welded joints that protrude up to approx. 5 mm (0.2") inwards
do not influence the measurement.
• The pipe may not exhibit any steps in diameter.
• The probe must be 100 mm longer than the lower disposal.
• Within the measuring range, the probe must not get into contact with the pipe wall. If
necessary, use a center washer (see feature 610 of the product structure).
• If the center washer is mounted at the end of the probe, it enables a reliable recognition of the
end-of-probe signal (see feature 610 of the product structure).
Note: For interface measurements only use the nonmetallic center washers made of PEEK or
PFA (feature 610, options OD or OE) (® ä 96).
• Coax probes can always be applied if there is enough mounting space.