EasyManua.ls Logo

Endress+Hauser Micropilot FMR56 - Language Selection

Endress+Hauser Micropilot FMR56
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR56, FMR57 FOUNDATION Fieldbus Commissioning (block-based operation)
Endress+Hauser 83
0
1
0
1
EU_100
XD_SCALE OUT_SCALE
EU_0
EU_0
EU_100
OUT_VALUE
A0017338
 23 Scaling of the measured value in an AI Block
If you have selected the Direct mode for the L_TYPE parameter, you cannot change
the values and units for XD_SCALE and OUT_SCALE.
The L_TYPE, XD_SCALE and OUT_SCALE parameters can only be changed in the
OOS block mode.
12.4 Language selection
Step Block Parameter Action
1 DISPLAY (TRDDISP) Language (language) Select language
1)
.
Selection:
1268: Swedish
32805: Arabian
32824: Chinese simplified
32842: Czech
32881: Dutch
32888: English
32917: French
32920: German
32945: Italian
32946: Japanese
32948: Korean
33026: Polish
33027: Portuguese
33062: Russian
33083: Spanish
33103: Thai
33120: Vietnamese
33155: Bahasa
33166: Turkish
1) When ordering a device the set of available languages is defined. Refer to the product structure, feature
500 "Additional Operation Language".