Operating menu Micropilot FMR56, FMR57 FOUNDATION Fieldbus
168 Endress+Hauser
"Configuration backup display" submenu
This submenu is only visible if a display module is connected to the device.
The configuration of the device can be saved to the display module at a certain point of
time (backup). The saved configurateion can be restored to the device if required, e.g. in
order to bring the device back into a defined state. The configuration can also be
transferred to a different device of the same type using the display module.
Navigation  Setup → Advanced setup → Conf.backup disp
Operating time
Navigation  Setup → Advanced setup → Conf.backup disp → Operating time
Description Indicates how long the device has been in operation.
User interface Days (d), hours (h), minutes (m), seconds (s)
Additional information Maximum time
9 999 d ( ≈ 27 years)
Last backup
Navigation  Setup → Advanced setup → Conf.backup disp → Last backup
Description Indicates when the last data backup was saved to the display module.
User interface Days (d), hours (h), minutes (m), seconds (s)
Configuration management
Navigation  Setup → Advanced setup → Conf.backup disp → Config. managem.
Description Select action for managing the device data in the display module.
Selection Cancel
Execute backup
Restore
Duplicate
Compare
Clear backup data
Display incompatible
Factory setting Cancel