Micropilot FMR56, FMR57 FOUNDATION Fieldbus Integration into a FOUNDATION Fieldbus network
Endress+Hauser 61
9 Integration into a FOUNDATION Fieldbus
network
9.1 Device Description (DD)
You require the following to configure a device and integrate it into an FF network:
An FF configuration program
The Cff file (Common File Format: *.cff)
The device description (DD) in one of the following formats
Device Description format 4 : *sym, *ffo
Device Description format 5 : *sy5, *ff5
Information on the device-specific DD
Manufacturer ID 0x452B48
Device Type 0x1028
Device Revision 0x01
DD Revision Information and files at:
www.endress.com
www.fieldbus.org
CFF Revision
9.2 Integration into the FOUNDATION Fieldbus network
For more in-depth information on integrating the device into the FF system, see the
description for the configuration software used.
When integrating the field devices into the FF system, make sure you are using the
right files. You can read out the required version by means of the Device Revision/
DEV_REV and DD Revision/ DD_REV parameters in the Resource Block.
The device is integrated into the FF network as follows:
1. Start the FF configuration program.
2. Download the Cff and device description files (*.ffo, *.sym (for format 4) *ff5, *sy5
(for format 5) to the system.
3. Configure the interface.
4. Configure the device for the measuring task and for the FF system.
9.3 Device identification and addressing
FOUNDATION Fieldbus identifies the device using its ID code (Device ID) and automatically
assigns it a suitable field address. The identity code cannot be changed. The device appears
in the network display once you have started the FF configuration program and integrated
the device into the network. The blocks available are displayed under the device name.
If the device description has not yet been loaded, the blocks report "Unknown" or "(UNK)".