Proline Promag P 100
28 Endress+Hauser
DN 300£ DN 350³
A0016317
 15 Potential equalization via ground terminal and pipe flanges
Note the following when installing:
Connect both sensor flanges to the pipe flange via a ground cable and ground them.
Connect the connection housing of the transmitter or sensor to ground potential by means of the
ground terminal provided for the purpose.To mount the ground cable:
If DN ≤ 300 (12"): Mount the ground cable directly on the conductive flange coating of the
sensor with the flange screws.
If DN ≥ 350 (14"): Mount the ground cable directly on the metal transport bracket.
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)
The necessary ground cable can be ordered from Endress+Hauser .
Plastic pipe or pipe with insulating liner
This connection method also applies in situations where:
The customary potential equalization is not used
Equalizing currents are present
A0016318
 16 Potential equalization via ground terminal and ground disks
Note the following when installing:
The ground disks must be connected to the ground terminal via the ground cable and be connected
to ground potential.
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)
The ground cable and ground disks can be ordered from Endress+Hauser (→  54).