Proline Promag P 100
Endress+Hauser 31
Maximum measured error Error limits under reference operating conditions
o.r. = of reading
Volume flow
±0.5 % o.r. ± 1 mm/s (0.04 in/s)
Optional: ±0.2 % o.r. ± 2 mm/s (0.08 in/s)
Fluctuations in the supply voltage do not have any effect within the specified range.
2.5
[%]
2.0
1.5
1.0
0.5
0
0.2 %
0.5 %
0
1
2 4 6 8 10 [m/s]
v
5 10 15 20 25 30 32 [ft/s]0
A0005531
 18 Maximum measured error in % o.r.
Electrical conductivity
Max. measured error not specified.
Accuracy of outputs
o.r. = of reading; o.f.s. = of full scale value
The output accuracy must be factored into the measured error if analog outputs are used, but
can be ignored for fieldbus outputs (e.g. Modbus RS485, EtherNet/IP).
Current output
Accuracy Max. ±0.05 % o.f.s. or ±5 µA
Pulse/frequency output
Accuracy Max. ±50 ppm o.r.
Repeatability
o.r. = of reading
Volume flow
Max. ±0.1 % o.r. ± 0.5 mm/s (0.02 in/s)
Electrical conductivity
Max. ±5 % o.r.
Temperature measurement
response time
T
90
< 15 s
Influence of ambient
temperature
o.r. = of reading; o.f.s. = of full scale value
Current output
Temperature coefficient Max. ±50 ppm/°C o.f.s. or ±1 µA/°C