Proline Promag P 100
46 Endress+Hauser
Nominal
diameter
Pressure rating Process connection internal diameter
EN
(DIN)
ASME AS
2129
AS
4087
JIS PFA PTFE
[mm] [in] [bar] [psi] [bar] [bar] [bar] [mm] [in] [mm] [in]
40 1 ½ PN 40 Class 150 20K 36 1.42 41 1.61
50 2 PN 40 Class 150 Table E PN 16 10K 48 1.89 52 2.05
65 PN 16 10K 63 2.48 67 2.64
80 3 PN 16 Class 150 10K 75 2.95 80 3.15
100 4 PN 16 Class 150 10K 101 3.98 104 4.09
125 PN 16 10K 126 4.96 129 5.08
150 6 PN 16 Class 150 10K 154 6.06 156 6.14
200 8 PN 10 Class 150 10K 201 7.91 202 7.95
250 10 PN 10 Class 150 10K 256 10.1
300 12 PN 10 Class 150 10K 306 12.0
350 14 PN 10 Class 150 337 13.3
400 16 PN 10 Class 150 387 15.2
450 18 PN 10 Class 150 432 17.0
500 20 PN 10 Class 150 487 19.2
600 24 PN 10 Class 150 23 593 23.3
Materials Transmitter housing
Order code for "Housing", option A "Compact, aluminum coated":
Coated aluminum AlSi10Mg
Cable entries/cable glands
1
2
3
A0020640
 28 Possible cable entries/cable glands
1 Cable entry in transmitter housing, wall-mount housing or connection housing with internal thread M20 x
1.5
2 Cable gland M20 x 1.5
3 Adapter for cable entry with internal thread G ½" or NPT ½"

Table of Contents