Proline Promag P 100
Endress+Hauser 49
A0016948
29 Options for remote operation via HART protocol
1 Control system (e.g. PLC)
2 Field Communicator 475
3 Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM)
4 Commubox FXA195 (USB)
5 Field Xpert SFX100
6 VIATOR Bluetooth modem with connecting cable
7 Transmitter
Via service interface (CDI-RJ45)
This communication interface is present in the following device version:
• Order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
• Order code for "Output", option L: PROFIBUS DP
• Order code for "Output", option N: EtherNet/IP
A0016926
30 Connection for the order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
1 Service interface (CDI -RJ45) of the measuring device with access to the integrated Web server
2 Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3 Standard Ethernet connecting cable with RJ45 plug