6 Commissioning Promag 53 PROFIBUS-DP/-PA
106 Endress+Hauser
6.5 Empty-pipe/full-pipe calibration
Flow cannot be measured correctly unless the measuring tube is full. The Empty Pipe
Detection function (EPD = Empty Pipe Detection) allows this status to be monitored con-
tinuously.
The EPD function cannot be activated until after empty/full pipe calibration has been
performed. This adjustment procedure is described below.
Note:
These settings can also be made in accordance with the manufacturer's specific
matrix using Commuwin II.
The EPD function is only available if the sensor is equipped with an EPD
electrode.
If the EPD electrode is installed, the sensor is calibrated ex-works with water, i.e. at
approximately 500
µS/cm. If the fluid conductivity differs from this reference, empty
pipe/full pipe calibration has to be performed again on site.
The default setting for EPD when the devices are delivered is OFF; the function has to
be activated if required.
The EPD process error is issued via the status and diagnosis messages.
Response to partially filled pipe
If EPD is on and trips in response to a partially filled or empty measuring tube, the mes-
sage PARTIALLY FILLED PIPE appears on the local display. The message is also
transferred in the status and diagnosis message to the control system during cyclic data
exchange
Page 115 ff.
If the pipe is partially filled or empty and EPD is not on, response can vary in identically
configured systems:
Flow reading fluctuates
Zero flow
Excessively high flow readings
Perform empty and full pipe calibration (EPD)
1. Select the appropriate function in the function matrix:
HOME
BASIC FUNCTION
PROCESS PARAMETER
ADJUSTMENT
EPD ADJUSTMENT
2. Empty the piping. Make sure that the wall of the measuring tube is wetted with fluid
for the calibration procedure.
3. Start empty pipe calibration:
Select EMPTY PIPE ADJUSTMENT setting and confirm with

.
4. Fill the pipe with fluid.
5. Start full pipe calibration with the fluid at a standstill:
Select FULL PIPE ADJUSTMENT setting and confirm with
.
6. Switch on Empty Pipe Detection after calibration
Select ON and confirm
with
.
Caution:
The calibration coefficients must be valid before you can activate the EPD function.
If calibration is incorrect the following messages might appear on the display:
EPD ADJUSTMENT FULL = EMPTY
The EPD calibration values for empty pipe and full pipe are identical.
EPD ADJUSTMENT NOT OK
EPD calibration is not possible because the fluid's conductivity is out of range.
In cases of this nature you must repeat empty-pipe or full-pipe calibration!
These messages will also be displayed in the cyclic status and in the diagnosis of
the control system (error messages
Page 115 ff.).