Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 113
9 Trouble-shooting
9.1 Trouble-shooting instructions
Always start trouble-shooting with the checklist below, if faults occur after start-up or
during operation. The routine takes you directly to the cause of the problem and the
appropriate remedial measures.
Caution:
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer
for repair. The procedures on Page 6 must be carried out before you return a device to
Endress+Hauser. Always enclose a duly completed Safety regulation form. You will
find a preprinted form at the back of this manual.
Check the display
No display visible and no
output signals present.
1. Check the power supply terminals 1, 2
2. Check device fuse
Page 127
85...260 V AC: 0,8 A slow-blow / 250 V
20...55 V AC and 16...62 V DC: 2 A slow-blow / 250 V
3. Measuring electronics defective
order spare parts Page 122
No display visible, but out-
put signals are present.
1. Check whether the ribbon-cable connector of the display module is
correctly plugged into the amplifier board
Page 124, 126
2. Display module defective
order spare parts Page 122
3. Measuring electronics defective
order spare parts Page 122
Display texts are in a foreign
language.
Switch off power supply. Press and hold down both the +/
buttons and
switch on the measuring device. The display text will appear in English
(default) and is displayed at maximum contrast.
Error messages on display
Errors which occur during commissioning or measuring operation are displayed immediately.
Error messages consist of a variety of icons. The meanings of these icons are as follows:
Error type: S = system error, P = process error
Error message type:
= fault message, ! = notice message
PARTIALLY FILLED PIPE = error designation (e.g. measuring tube is only partly filled)
03:00:05 = duration of error occurrence (in hours, minutes and seconds)
#401 = error number
Caution:
See the information on Page 69 ff.!
Simulations and positive zero return are interpreted by the measuring system as being system errors-
but are only displayed as notice messages.
Error number:
No. 001 111
No. 205 321
No. 601 692
System error (device failure) has occurred
Page 115 ff.
Error number:
Nr. 401
463
Process error (application error) has occurred
Page 115 ff.