Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 127
9.6 Replacing the device fuse
Warning:
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that
the power supply is switched off before you remove the cover of the electronics com-
partment.
The main fuse is on the power supply board (Fig. 56).
The procedure for replacing the fuse is as follows:
1. Switch off power supply.
2. Remove the power supply board
Page 123, 125
3. Remove cap (1) and replace the device fuse (2).
Use only fuses of the following type:
Power supply 20...55 V AC / 16...62 V DC
2.0 A slow-blow /
250 V; 5.2 x 20 mm
Power supply 85...260 V AC
0.8 A slow-blow / 250 V; 5.2 x 20 mm
Ex-rated devices
see the Ex documentation.
4. Assembly is the reverse of the disassembly procedure.
Caution:
Use only original Endress+Hauser parts.
Fig. 56: Replacing the device fuse on the power supply board
1 Protective cap
2 Device fuse