Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 61
4.5 Degree of protection
The devices fulfill all the requirements for IP 67. Compliance with the following points is
mandatory following installation in the field or servicing, in order to ensure that IP 67 pro-
tection is maintained:
The housing seals must be clean and undamaged when inserted into their grooves.
The seals must be dried, cleaned or replaced if necessary.
All threaded fasteners and screw covers must be firmly tightened.
The cables used for connection must be of the specified outside diameter
(see Page 132).
Firmly tighten the cable entries (Fig. 40).
The cables must loop down before they enter the cable entries (water trap, Fig. 40).
This arrangement prevents moisture penetrating the entry. Always install the measur-
ing device in such a way that the cable entries do not point up.
Remove all unused cable entries and insert plugs instead.
Do not remove the grommet from the cable entry.
Fig. 40: Installation instructions, cable entries
Caution:
Do not loosen the threaded fasteners of the Promag sensor housing, as otherwise the
degree of protection guaranteed by Endress+Hauser no longer applies.
Note:
The Promag W and Promag P sensors can be supplied with IP 68 rating (permanent
immersion in water to a depth of 3 meters). In this case the transmitter must be installed
remote from the sensor.