Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 41
3.3.5 Installing the wall-mount transmitter housing
There are various ways of installing the wall-mount transmitter housing:
Mounted directly on the wall (without mounting kit)
Installation in control panel (with separate mounting kit, accessories Page 111)
Pipe mounting (with separate mounting kit, accessories Page 111)
Caution:
Make sure that ambient temperature does not exceed the permissible range
(20 °...+60 °C). Install the device at a shady location. Avoid direct sunlight.
Always install the wall-mount housing in such a way that the cable entries are pointing
down.
Direct wall mounting
1. Drill the holes as illustrated in Fig. 25.
2. Remove the cover of the connection compartment (a).
3. Push the two securing screws (b) through the appropriate bores (c) in the housing.
Securing screws (M6): max. Ø 6.5 mm
Screw head: max. Ø 10.5 mm
4. Secure the transmitter housing to the wall as indicated.
5. Screw the cover of the connection compartment (a) firmly onto the housing.
Fig. 25: Mounted directly on the wall
F06-xxxxxxxx-17-03-xx-xx-000