3 Installation Promag 53 PROFIBUS-DP/-PA
42 Endress+Hauser
Panel installation
1. Prepare the opening in the panel.
2. Slide the housing into the opening in the panel from the front.
3. Screw the fasteners onto the wall-mount housing.
4. Place the threaded rods in the fasteners and screw them down until the housing is
seated tightly against the panel. Afterwards, tighten the locking nuts. Additional
support is not necessary.
Abb. 26: Panel Installation (wall-mount housing)
Pipe mounting
The assembly should be performed by following the instructions in Fig. 27.
Caution:
If the device is mounted to a warm pipe, make certain that the ambient temperature
does not exceed +60 °C, which is the maximum permissible temperature.
Fig. 27: Pipe mounting (wall-mount housing)
245
~110
+0.5
0.5
210
+0.5
0.5
F06-xxxxxxxx-06-03-06-xx-002
Ø 20...70
~155
F06-xxxxxxxx-06-03-06-xx-001