Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 25
3.3 Installation
3.3.1 Installing the Promag W sensor
Note:
Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by
the customer.
The sensor is designed for installation between the two piping flanges. Always tighten
all threaded fasteners to the specified torques
Page 27 ff.
Fig. 16: Installing the Promag W sensor
Seals
Comply with the following instructions when installing seals:
Hard rubber lining additional seals are always necessary!
Polyurethane lining additional seals are recommended
For DIN flanges, use only seals acc. to DIN 2690.
Make sure that the seals do not protrude into the piping cross-section.
Caution:
Risk of short circuit. Do not use electrically conductive sealing compound such as
graphite. An electrically conductive layer could form on the inside of the measuring tube
and short-circuit the measuring signal.
Grounding cable (DN 15...2000)
If necessary, the special grounding cable for potential equalisation can be ordered as
an accessory (see Page 111). Detailled assembly instructions
Page 59 ff.
F06-5xFxxxxx-17-05-xx-xx-000