Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 119
EPD adjustment wrong P EPD FULL=EMPTY
# 463
463 non specific UNCERTAIN
0x40
no limits
Cause:
Process error. The EPD calibration
values for empty pipe and full pipe
are identical, therefore incorrect.
Remedy:
Repeat EPD calibration, making
sure procedure is correct
Page 106
Positive zero return
active
S POS.ZERO-RET.
! # 601
601 sensor conversion not
accurate (measured
value from sensor not
accurate)
UNCERTAIN
0x53
constant
Cause:
System error
Positive zero return is activated
Remedy:
Switch off positive zero return
PROFIBUS (Commuwin II):
Manufacturer-specific trans-
ducer block
Device matrix
(V8H5)
Local display:
BASIC FUNCTION
SYSTEM
PARAMETER
CONFIGURA-
TION
POSITIVE ZERO
RETURN (
OFF)
Simulation failsafe
active
S SIM. FAILSAFE
! # 691
691 substitute set UNCERTAIN
0x48...0x4B
low/high
constant
Cause:
System error
Simulation of response to error is
active
Remedy:
Switch off simulation
PROFIBUS (Commuwin II):
Manufacturer-specific trans-
ducer block
Service & Analy-
sis (V4H2)
Local display:
SUPERVISION
SYSTEM
OPERATION
SIM. FAILSAFE
MODE (
OFF)
Simulation Volume flow
active
S SIM. VOL. FLOW
! # 692
692 simulated value
(manually defined
value)
UNCERTAIN
0x60...0x63
low/high
constant
Cause:
System error
Simulation is active
Remedy:
Switch off simulation
PROFIBUS (Commuwin II):
Manufacturer-specific trans-
ducer block
Service & Analy-
sis (V4H0)
Local display:
SUPERVISION
SYSTEM
OPERATION
SIM. MEASU-
RED VARIABLE (
OFF)
Device status
message
Diagnosis message
(Control system)
Device status
message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Q /
Substatus/
alarm limit
Cause/remedy