EasyManua.ls Logo

Endress+Hauser Promag 53 - Page 129

Endress+Hauser Promag 53
177 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 129
Removing the electrode Installing the electrode
1 Loosen Allen screw (a) and remove the
cover.
1 Insert new electrode (e) into retaining cylin-
der (g) from below. Make sure that the seals
at the tip of the electrode are clean.
2 Remove electrode cable (c) secured to
handle (b).
2 Mount handle (b) on the electrode and
insert locking pin (h) to secure it in position.
Caution:
Make sure that coil spring (k) is inserted.
This is essential to ensure correct electrical
contact and correct measuring signals.
2 Loosen knurled nut (d) by hand. This
knurled nut acts as a locknut.
2 Pull the electrode back until the tip of the
electrode no longer protrudes from retain-
ing cylinder (g).
3 Remove electrode (e) by turning handle
(b). The electrode can now be pulled out of
retaining cylinder (g) as far as a defined
stop.
Warning:
Risk of injury! Under process conditions
(pressure in the piping system) the elec-
trode can recoil suddenly against its stop.
Apply counter-pressure while releasing the
electrode.
3 Screw the retaining cylinder (g) onto ball-
valve housing (i) and tighten it by hand.
Seal (j) on the cylinder must be correctly
seated and clean.
Note:
Make sure that the rubber hoses on retain-
ing cylinder (g) and stop cock (f) are of the
same color (red or blue).
4 Close stop cock (f) after pulling out the
electrode as far as it will go.
Warning:
Do not subsequently open the stop cock,
in order to prevent fluid escaping.
4 Open stop cock (f) and turn handle (b) to
screw the electrode all the way into the
retaining cylinder.
5 Remove the electrode complete with
retaining cylinder (g).
5 Screw knurled nut (d) onto the retaining cyl-
inder. This firmly locates the electrode in
position.
6 Remove handle (b) from electrode (e) by
pressing out locking pin (h). Take care not
to lose coil spring (k).
6 Use the Allen screw to secure electrode
cable (c) to handle (b).
Caution:
Make sure that the machine screw securing
the electrode cable is firmly tightened. This
is essential to ensure correct electrical con-
tact and correct measuring signals.
7 Remove the old electrode and insert the
new electrode. Replacement electrodes
can be ordered separately from E+H.
7 Reinstall the cover and tighten (a) Allen
screw.

Table of Contents

Related product manuals