Wiring TOCII CA72TOC
26 Endress+Hauser
4.2 Post-connection check
Carry out the following checks after electrical connection:
Device status and specifications Note
Is the exterior of the analyzer or cable undamaged? Visual inspection
Electrical connection Note
Does the supply voltage correspond to the data on the nameplate? 230 V AC 50/60 Hz
115 V AC 50/60 Hz
Are current outputs shielded and connected?
Are the mounted cables strain-relieved?
Are the cable types properly separated? Run power supply and signal lines
separately over the entire travel distance.
Separate cable channels are ideal.
Are the cables routed without loops or crossovers?
Are power supply and signal lines connected correctly according to wiring
diagram?
Are all screw terminals tightened?
Are all cable entries mounted, tightened and leak-resistant?